Download Brochure
The PULSI
200LR series of instruments are designed for
in-stack analysis of up to six gas phase stack
emission components.
The PULSI 200LR system comprises four standard
units: the PULSI 200LR OPTICAL HEAD UNIT (OHU),
the ANALYSER CONTROL UNIT (ACU), and the
AUTOZERO UNIT (AZU). An IN-SITU HEATER (ISH) is
optionally supplied when process conditions are
liable to result in water condensing in the
analyser. The instrument operates on the dual
wavelength infra-red principle where pulses of
two wavelengths, per gas, at specific
wavelengths, are sent through the sample cell of
the STACK MOUNTED OPTICAL HEAD UNIT (OHU). . A total of 8
wavelengths are available and in some
circumstances reference wavelengths are shared.
This allows up to six (6) gas phase
concentrations to be measured simultaneously.
The specific wavelengths used for analysis are
application dependent and selected by the Procal
Analytics Application Group before each
instrument is calibrated. The selected infra-red
wavelengths (8) used for analysis are obtained
by means of interference filters and gas filled
cells (GFC) mounted on a rotating wheel located
in the STACK MOUNTED OPTICAL HEAD UNIT. The
PULSI 200LR Series has resulted from the
culmination of many years research and
development into stack gas analysis. The
advanced stack gas analyser utilises an in situ
(inside the stack) sample cell thereby avoiding
the need to extract a sample from the stack.
This avoids the use of costly, high maintenance
sample handling systems, but more importantly
analyses an UNMODIFIED, truly representative gas
sample. Sintered stainless steel filter panels
fitted to the sides of the in-stack measuring
cell allow the permeation of stack gas whilst
preventing the ingress of dust and particulates.
The envelope thus formed allows the introduction
of zero and span gases. The OPTICAL HEAD UNIT
requires no independent power supply; all
necessary power is supplied via the ACU
interconnection cable. The ANALYSER CONTROL UNIT
(ACU) displays gas concentrations on the
integral liquid crystal display along with
information on sample conditions, diagnostic
data and trends. The information can be
retransmitted in the form of 4-20mA current
outputs (one per measured component), parallel
printer interface and optional RS232C serial
output. The OHU controls the AUTOZERO UNIT. On
command from the ANALYSER CONTROL UNIT, the AZU
will initiate a zero check on the system by
filling the sample probe with zero gas typically
clean, dry instrument air. In addition to the
auto zero function, the AZU has a safety feature
which, in the event of power loss to the
analyser, or too low a sample temperature will
purge the sample cell with instrument air. This
is done to prevent highly corrosive condensates
forming in the sample cell. The OHU also
controls the temperature of the optional IN-SITU
HEATER which is used if the analyser is required
to operate in stacks or ducts where the sample
gas temperature is below its dew point. In this
application the ISH requires a mains supply.